Model: CD-36D 7.2 Yearly Consumable and Spare Parts / 7.3 Cleaning of Detector and Replacement of Sensing Electrode
-
24
-
7.2 Yearly Consumable and Spare Parts
The following table shows general consumables and spare parts. Be prepared with them in your stock
just in case, and please note that they vary with the specification of the Transmitter, measuring
conditions, etc. and are subject to change without notice.
Yearly Consumable and Spare Parts
Quantity
No. Code No. Item
Sketch
Consumables
Spare
Parts
Remarks
1 123G007 Ceramic Beads
1 φ 1, 25g
For CLV Series and
CLR-21-A
2 123G008 Glass Beads
1 φ 1, 25g
For CLV Series
3 EL2132 Sensing Electrode
1 Model: 2132
For CLR-21-A
4 104A072 Tubular Fuse
2 For Transmitter
5 519052K Electrically conductive
grease assembly
1 For CLR-21-A, 10g
7.3
Cleaning of
Detector
and Replacement of Sensing Electrode
① Prepare washing solution. ⋅⋅⋅⋅⋅⋅ Prepare 300mL of city water in a 500mL beaker and add ca. 1mL
of neutral detergent into it.
② Turn off the power switch. ⋅⋅⋅⋅⋅⋅ Open the rear cap of the Transmitter and turn off the power switch
located on the upper part of the terminal board (Power: OFF).
③ Disconnect the lead wires. ⋅⋅⋅⋅⋅⋅ Turn the closing ring for the connector on the upper part of the
detector counterclockwise and disconnect the connector for the lead wires.
<In the case of Model: CLR-21-A>
O-ring
Flow Cell
Lead Wires
Connector
Detector
Cap
Sensing
Electrode
Closing Ring
<In the case of CLV Series>
O-ring
Flow Cell
Lead Wires
Connector
Detector
Rotate to
remove.
Cap
Removal of the detector