3-4
3.2.3 Clearing the last examination data (preparation for a new patient)
ڦ
The deleted examination data cannot be restored. Carefully check the data
before deleting it.
ڦ
Be sure to touch the “New” button to delete the examination data for the
previous patient before capturing images for a new patient. Otherwise, the
data for the previous patient may be included.
ڦ
This operation does not clear examination data saved in the database. Refer
to “3.6.2 Deleting examination data” to delete the examination data in the
database.
1) Touch the “New” button for a while.
- The patient information (ID, name, sex, and date of birth) and relevant
examination data is deleted, and the capture screen appears.
(Fig. 1)
3.2.4 Entering the patient data
ڦ
When entering new patient information, carefully check that the same ID is
not registered in the database or other electronic medical records.
ڦ
Only the first 14 digits of the ID number are displayed in the patient
information field. Check that the ID number is correct on the patient
information entry screen.
ڦ
The network support system “TOMEY Link” (optional) is required to use the
patient information query function. “DATA Transfer” provided with the
Machine is not available for query of the patient data.
ڦ
For details on the TOMEY Link Server settings, refer to the TOMEY Link
instruction manual.
ڦ
Connection setting on the System Setup screen must be completed before
connecting with TOMEY Link. Refer to “3.7.4 Connection & Print” for the
setting method.
ڦ
Once examination data is output (printed, saved, or exported), the ID cannot
be changed.
1) Touch the patient information button (1) to open the Patient Information
screen (Fig. 2).
- When both the patient information and examination data exist, a
Warning (Fig. 3) appears indicating you are about to overwrite data
when the patient information button is touched. Touch the “Yes” button
(7) to overwrite the data.