䕔
7-2
7.4 Declaration of Conformity to EMC
The EM-4000 belongs to Group 1 and B equipment in accordance with
IEC/EN60601-1-2.
The EM-4000 is a group 1, class B product according to
EN55011(CISPR11). This means that the EM-4000 does not
generate and/or use intentionally ratio-frequency energy, in the
form of electromagnetic radiation, inductive and/or capacitive
coupling, for the treatment of material or inspection/analysis
purpose and that it is suitable for use in domestic establishments
and in establishments directly connected to a low voltage power
supply network which supplies buildings used for domestic
purposes.
The EM-4000 requires special precautions regarding EMC and need to
be installed, put into service and used according to the following
information.
Do not use any cables other than the cables that provided or specified by
Tomey.
The use of accessories, transducers and cables other than those
specified, with the exception of transducers and cables sold by Tomey
as replacement parts for internal components, may result in increased
EMMISSIONS or decreased IMMUNITY of the EM-4000.
Portable and mobile RF communications equipment can affect the EM-4000.
The EM-4000 should not be used adjacent to or stacked with
other equipment and if adjacent or stacked use is necessary, the
EM-4000 should be observed to verify normal operation in the
configuration in which it will be used.
Anyone who connects additional equipment to the signal
input/output parts, configuration a medical system, is
responsible that the system complies with the requirements of
IEC/EN60601-1-2.
z Cable List
No. cable Maximum length [m] Remarks
1 AC code 2.5 Unshielded
z Accessory List
No. name model manufacture
1 AC code E32-0025 KDK EKECTRIC WIRE