3-41
(10) ID INPUT
Set whether to display the patient information input screen before
capturing images when starting to capture images for a new patient.
When “(9) ID Necessity, SAVING / EXPORT” setting is “No”, this setting
is automatically set to “No” and cannot be selected.
Yes : Opens the patient information entry screen before
capturing images.
No : Opens the capture screen.
(11) Input Mask
Sets which part of the character string should be recognized as the ID
number when entering the ID number from a barcode. If it starts from “0”
or “1”, the characters from the very beginning are identified as the ID
number.
(12) Confirmation
Select whether to open the patient information confirmation screen
when saving or sending the examination data to TOMEY Link.
3.7.2 Measurement
Make settings related to capturing images. Select the “Measurement” tab.
(Fig. 1)
(1) Measurement mode
Sets the mode for capturing images.
Quick : Capture mode in which the image selection screen does
not appear when capturing is completed. This mode is for
users that do not need to look at all 16 images of the
endothelium tissues.
Basic : Capture mode in which the image selection screen
appears when capturing is completed. This mode is for
users who want to look at all 16 captured images of the
endothelium tissues and select only one from them.
(2) Joystick
Activates or deactivates the joystick.
(1)
(2)