3-15
3.4 Analysis
3.4.1 Auto Analysis
Auto Analysis is to calculate values such as CD (cell density) by automatically extracting trace lines
of cells on the captured image
ڦ
Check that the trace line of cells is correctly detected on the captured image.
The trace line cannot be extracted on some images.
ڦ
Perform any of the following operations if the trace line is not extracted
correctly.
- Retake
ure an image again if the images are not clear.
- Correcting trace line
Correct and redraw the incorrect images.
Refer to “3.4.1 Editing trace line” for how to edit the trace line..
- Selecting cell
Exclude cells with trace lines extracted incorrectly from the target of analysis.
Refer to “3.4.1 c) Selecting cells” for how to select cells.
- Changing analysis method
Perform analysis with the L-count method or Core method. Refer to “3.4.2 L-count
method” and “3.4.3 Core method” for the operation of the L-count method and Core
method.
a) How to recognize capacity data
Ɣ R/L eye fracture screen
When touching the “Analyze” button on the capture screen or image
selection screen, the captured image is analyzed automatically and the R/L
eye analysis screen (Fig. 1) appears. In Quick mode, the image selection
screen appears when images of both eyes have been captured.
(Fig. 1)
(1)