1-1
1. PRIOR TO USE
ڦ
Read this manual thoroughly before using the instrument to ensure
proper and safe operation.
ڦ
Always follow the operation procedures described in this manual.
ڦ
Check that there are no devices that generate strong magnetic field near
the instrument. A strong magnetic field may cause noise and affect the
capture of images.
1.1 Precautions for operation
Only allow qualified operators to use the instrument.
Precautions when installing the instrument
Install the instrument in a location free of water or chemicals. Any water
or chemicals entering the instrument may cause an electric shock or
failure.
Do not install the instrument in a location where chemicals are stored or
gases may occur. Spilt chemicals or vapor may enter the instrument and
result in fire.
Check that the frequency, voltage, and allowable current (or power
consumption) of the power source are appropriate. Otherwise, fire or
electric shock may occur.
Connect the power plug to a grounded 3-pin outlet. Otherwise, a short
circuit due to failure of the instrument may result in electric shock.
Do not place any heavy object on the power cord or squash the power
cord. Fire or electric shock may occur.
Fully insert the power plug into the outlet. Faulty contact, allowing any
metal to contact the exposed terminal of the plug, or dust accumulated
on the exposed terminal of the plug may result in fire or electric shock.
When operating this instrument connected to other devices, only use
devices that comply with IEC60601-1 or that comply with IEC60950-1
and whose power source is isolated with an isolation transformer.
Furthermore, all devices should be configured to comply with the
standard IEC60601-1 ME system. Anyone who connects any
additional device to the communication connector will be considered a
person configuring a medical system, and is therefore responsible for
complying with the requirements of an IEC60601-1 ME system.
Contact us or our local distributor before connecting the instrument to
another device with a communication connector.
Do not connect a device with data transmission specifications that are
not compatible. Fire or electric shock may occur. Contact us or our local
distributor before using the instrument while connected to another
device.
Conduct grounding work correctly. Otherwise you may get an electric
shock.