z Basic performance
࣭Corneal endothelial cell density calculated by analysis of captured image is
indicated properly.
࣭Measurement data of central corneal thickness is indicated properly.
Caution: Medical electrical equipment.
EMC (Electro Magnetic Compatibility) must be considered before any medical electrical
equipment is installed or put into service. Follow the information in the accompanying
documentation when installing and operating the EM-4000.
Caution: Portable or mobile RF communication equipment can effect Medical Electrical
equipment.
Guidance and manufacturer's declaration -
electromagnetic emissions
The EM-4000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the EM-4000 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The EM-4000 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class B The EM-4000 is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuation/
flicker emissions
IEC 61000-3-3
Complies
Table 201