5-3䢢
䕔
The printer cover is open.
Check that the printer cover is completely closed.
The printer paper is not set in the correct direction.
Check that the printer paper is set correctly. Refer to
“4.5.2 Built-in printer paper” for the correct setting.
z A different button to the one touched on the touch panel
becomes active.
The touch panel is not calibrated correctly.
Contact our local distributor.
z Nothing is displayed on the screen of this instrument when
inputting IDs from the external ID input device.
USB cable.
Verify that the USB cable of the external ID input
instrument is correctly connected to the main unit.
Refer to “3.2.1 Connections" for correct connection.
A screen which does not receive an ID input appears.
Open the screen that receives the external ID input
device and then input IDs.
Refer to “3.2.4 Entering the patient data” for the
screen that receives the external ID input device.
z Data cannot be saved in the database.
The ID is not entered.
The instrument uses the ID as a file name. Enter the
ID and then save the data.
There is not enough space available in the SD card.
(The confirmation screen appears when attempting to
save the data.)
Delete unnecessary data in the SD card and save
the data again.