Details are as follows:
R Set “function” as R.
V Set “function” as V.
RV Set “function” as R-V.
RQ Set “function” as R-Q.
LQ Set “function” as L-Q.
LR Set “function” as L-R.
RX Set “function” as R-X.
ZTD Set “function” as Z-θ°.
ZTR Set “function” as Z-θr.
CD Set “function” as C-D.
For example: WrtCmd(“FUNC:IMP RX”); Set “function” as R-X .
Query syntax: FUNCtion:IMPedance?
Return Format: <function><NL^END>
The :IMPedance:RANGe command sets the instrument range.
The FUNCtion:IMPedance:RANGe? query returns the current range parameters.
Command Syntax: FUNCtion:IMPedance:RANGe <value>
Where, <value> is the positive integer from 0~5.
0 30mΩ
1 300mΩ
2 3Ω
3 30Ω
4 300Ω
5 3kΩ
For example: WrtCmd(“FUNC:IMP:RANG 0”); Set the instrument range as 30mΩ .
Query syntax: FUNCtion:IMPedance:RANGe?
Return format: <value><NL^END>
Where, <value> can be:
3m 300m 30
30 300 3k