43
SPECIFICATIONS AND PERFORMANCE
SYSTEM CLASSIFICATION
Type of protection against electric shocks: Class Ι equipment
Class Ι equipment does not depend on basic insulation only for protection against electric
shocks. It can also be earthed; therefore, the metal parts with which one comes into contact
do not become conductive if the basic insulation fails.
Degree of protection against electric shocks: Type B applied part
Type B applied part is the applied part complying with the specified requirements of the
Standard IEC 60601-1 to provide protection against electric shock, particularly regarding
allowable LEAKAGE CURRENT.
Degree of protection against harmful ingress of water: IPx0
SL-D7/D8Z has no protection against ingress of water. (The degree of protection against
harmful ingress of water defined in IEC 60529 is IPx0)
Classification according to the methods of sterilization or disinfection recommended by the
manufacture: not applicable.
SL-D7/D8Z has no part to be sterilized or be disinfected.
Not AP or APG equipment
Classification according to the degree of safety of application in the presence of a flamma-
ble anaesthetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for
use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous
oxide.
SL-D7/D8Z should be used in environments where no flammable anesthetics and/or flam-
mable gases are presents.
Classification according to the mode of operation: Continuous operation.
Continuos operation is the operation under normal load for an unlimited period, without
the specified limits of temperature being exceeded.
PURPOSES OF USE
This instrument is used for the enlargement observation of eyeballs and the parts.
OPERATION PRINCIPLES
Illuminates the observed part by the illumination light emitted from the illumination optical
system and allows enlargement observation by binocular stereoscopic microscope.

Table of Contents

Other manuals for Topcon SL-D7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon SL-D7 and is the answer not in the manual?

Topcon SL-D7 Specifications

General IconGeneral
BrandTopcon
ModelSL-D7
CategoryMedical Equipment
LanguageEnglish

Related product manuals