EasyManuals Logo

Topcon SL-D7 User Manual

Topcon SL-D7
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
51
MAINTENANCE AND CHECKUPS
DAILY CARE
CLEANING APPLIED PARTS
Wipe the forehead rest, the chinrest and the patient grip (if a pair of patient grip is used) with a
cloth moistened with a tepid solution of neutral detergent for kitchenware.
CLEANING LENSES AND MIRRORS
REMOVING STAINS
1 Prepare a solution of ethyl alcohol 20% and ether 80%.
2 Remove dust from lens and mirror surfaces with the cleaning brush, or a blower.
3 Using clean gauze or lint-free tissue, lightly clean with a rotating movement from the
center of the lens/mirror outwards.
4 If the stain remains, repeat this 2 to 3 times.
5 If stains are persistent, call your dealer or TOPCON (see the back cover).
CAUTION
Before carrying out daily care, remove the power cable (to
avoid electric shocks) and wait until the lamp house has cooled
(to avoid burns).
CAUTION
Do not touch parts inside the lamp house cover during opera-
tion and immediately after switching off the power supply: this
could cause burns.
NOTICE
To prevent the chinrest, forehead rest and other plastic
parts from discoloration and deterioration, do not use volatile
solvents for cleaning, including benzine, thinner, ether,
gasoline, etc.
Wipe parts with a cloth moistened with a tepid solution of
neutral kitchen detergent.
NOTICE
To prevent damaging lens surfaces, do not hold gauze with
tweezers.

Table of Contents

Other manuals for Topcon SL-D7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon SL-D7 and is the answer not in the manual?

Topcon SL-D7 Specifications

General IconGeneral
BrandTopcon
ModelSL-D7
CategoryMedical Equipment
LanguageEnglish

Related product manuals