60
SPECIFICATIONS & PERFORMANCE
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC Standard(IEC 60601-1-2:2001).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS.
b)Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIP-
MENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the exception
of transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replace-
ment parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY
of the EQUIPMENT or SYSTEM.
d)The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment. IF
adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify
normal operation in the configuration in which it will be used.
e)The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than
those specified may result in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or
SYSTEM.
Item Article code Model No. Length(m)
VIDEO OUT CABLE(BNC) - - 3.0
RS-232C CROSSING CABLE(shielded) - - 3.0
USB CABLE (USB 2.0) 419030201 - 3.0
USB MOUSE 419032700 0.65
CABLE VS200-SP 445102700 - 3.0
Guidance and manufacturer's declaration - electromagnetic emissions
The SP-3000P is intended for use in the electromagnetic environment specified below.
The customer or the user of the SP-3000P should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The SP-3000P uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The SP-3000P is suitable for use in all establishments,
excluding domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC61000-3-2
Does not
comply.
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Does not
comply.