SPECIFICATIONS & PERFORMANCE
ELECTRIC RATING
Source voltage : AC100V-240V, 50Hz/60Hz
Power input : 65VA for normal use, 190VA at its maximum
SYSTEM CLASSIFICATION
Types of protection against electric shocks:
This instrument is classified as Class I equipment.
Class I equipment does not depend only on basic insulation for protection against electric shocks,
but also provides a means of connection to a protective grounding system so that metal parts that
come into contact do not become conductive while the basic insulation is in failure.
Grade of protection against electric shocks:
This instrument is classified as Type B equipment.
Type B equipment provides a specified grade of protection to prevent electric shocks, particularly for
reliability against current leaks, measuring current and protective earth current (in case of Class I
equipment).
Degree of protection against harmful ingress of water: IPx0
The SP-3000P has no protection against ingress of water. (The degree of protection against harmful
ingress of water defined in IEC 60529 is IPx0.)
Classification according to the method(s) of sterilization or disinfection recommended by the manu-
facturer: not applicable.
The SP-3000P has no part to be sterilized or disinfected.
Classification according to the degree of safety of application in the presence of a flammable anes-
thetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in the presence
of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
The SP-3000P should be used in environments where no flammable anesthetics and/or flammable
gases are present.
Classification according to the mode of operation: Continuous operation.
Continuous operation is the operation under normal load for an unlimited period, without the speci-
fied limits of temperature being exceeded.
DIMENSIONS AND WEIGHT
Dimensions : 275mm (W) × 507mm (D) × 446~474mm (H)
Weight : 22kg
PURPOSE OF USE
Used to photograph and record the corneal endothelium as an electronic image.