10 AU-OPR-400Ser-EN,
Rev D
ii. TS410 Module: Specications
GENERAL FEATURES
Size: 5.125” h x 4” w x 9.062” d
Weight: 2.3 lbs.
Module ts 2 Console bays (20HP) in T402 or T403
Consoles
Power: Derives input power from 400-Series
Consoles. Installation in a Console is required.
RoHS compliant
OPERATIONAL TECHNOLOGY
Ultrasonic Transit-time
FLOWSENSOR COMPATIBILITY
ME-PXL-Series & ME-PXN-Series
SENSOR CONNECTOR
Front panel 16-pin connector. Accepts research
Inline and Clamp-on Flowsensors and extension
cables with male CC16 or CP16 connectors.
AUTOMATIC ADJUSTMENTS
Sensor size identication and corresponding ow
output ranges. Volume ow calibration and serial
number displayed of active Flowsensor.
LED DIGITAL DISPLAY
4-Digit (14 segment) LED displays Flow / Sensor
data / Error Messages
Bar Indicator Light: Displays received signal for
continuous monitoring of Sensor signal quality.
LCD DISPLAY
One line 16-character alpha numeric LCD displays
program parameters, Sensor and Module status,
alarm settings. Default displays Sensor serial
number.
SET-UP/STATUS & PROGRAM PARAMETERS
STATUS MODE: (White labels) Status message
displayed on LCD.
Sensor Status: Sensor type & calibration
Meter Status: Active Flow Module settings &
alarm status
Alarm Mute: Audible alarm on/off toggle
PROGRAM MODE: (Blue labels)
Sensor Controls: Select pre-programmed factory
calibration options; Adjust Flowsensor gain to
change calibration on-site.
1/4 Flow Scale: Increases ow gain by factor of 4
for low ow measurements.
Calibrate Scale: Sets output to 0 and 1 Volt to
calibrate external recording devices with scale
factor ow.
Invert Flow: Inverts polarity of analog outputs &
ow display
Alarms Menu: 3 level program to select, set
thresholds, and activate Alarms for “Low Flow”,
“High Flow” and “Received Signal” interruption
FILTER PROPERTIES
0.1, 10, 40 Hz: 2nd order Butterworth, with a
third passive pole at 160 Hz
160 Hz: 3rd order Butterworth
ZERO FLOW ADJUST
Recessed momentary push button to zero ow
reading at stopped ow.
FLOW OUTPUT
Front panel BNC output connector & rear panel
terminal block:
Pulsatile/Average Volume Flow
Filtering controlled by front panel selectable
lters
Voltage range: -5 to + 5 volts
Output resistance: 500 Ohm
Full Range for Flow: -5 to +5 V (bidirectional
ows, with range of 5 times scale factor ow)
AUTOMATIC DIGITAL SENSOR ID & CALIBRATION
TS410 reads operational data (size, scale &
calibration) programmed in the Sensor’s EPROM.
ULTRASONIC FREQUENCY RANGE
600 KHz to 14.4 MHz; Sensor size dependent
SIGNAL OUTPUTS
8 accessible signals via 400-Series Console’s back-
panel terminal block: Pulsatile Volume Flow; Mean
Volume Flow; Received Signal Amplitude (2);
Phase (4)
SYNCHRONIZATION
Rear panel jumpers select synchronization mode
Self-Triggering Mode: “SYNC IN” to “SYNC
OUT” jumper on each Module
Sequential Triggering Mode: “SYNC IN” crossed
to “SYNC OUT” between Modules