10 AU-OPR-400Ser-EN,
Rev D
ii. TS410 Module: Specications
GENERAL FEATURES
Size: 5.125” h x 4” w x 9.062” d
Weight: 2.3 lbs.
Module ts 2 Console bays (20HP) in T402 or T403
Consoles
Power: Derives input power from 400-Series
Consoles. Installation in a Console is required.
RoHS compliant
OPERATIONAL TECHNOLOGY
Ultrasonic Transit-time
FLOWSENSOR COMPATIBILITY
ME-PXL-Series & ME-PXN-Series
SENSOR CONNECTOR
Front panel 16-pin connector. Accepts research
Inline and Clamp-on Flowsensors and extension
cables with male CC16 or CP16 connectors.
AUTOMATIC ADJUSTMENTS
Sensor size identication and corresponding ow
output ranges. Volume ow calibration and serial
number displayed of active Flowsensor.
LED DIGITAL DISPLAY
4-Digit (14 segment) LED displays Flow / Sensor
data / Error Messages
Bar Indicator Light: Displays received signal for
continuous monitoring of Sensor signal quality.
LCD DISPLAY
One line 16-character alpha numeric LCD displays
program parameters, Sensor and Module status,
alarm settings. Default displays Sensor serial
number.
SET-UP/STATUS & PROGRAM PARAMETERS
STATUS MODE: (White labels) Status message
displayed on LCD.
● Sensor Status: Sensor type & calibration
● Meter Status: Active Flow Module settings &
alarm status
● Alarm Mute: Audible alarm on/off toggle
PROGRAM MODE: (Blue labels)
● Sensor Controls: Select pre-programmed factory
calibration options; Adjust Flowsensor gain to
change calibration on-site.
● 1/4 Flow Scale: Increases ow gain by factor of 4
for low ow measurements.
● Calibrate Scale: Sets output to 0 and 1 Volt to
calibrate external recording devices with scale
factor ow.
● Invert Flow: Inverts polarity of analog outputs &
ow display
● Alarms Menu: 3 level program to select, set
thresholds, and activate Alarms for “Low Flow”,
“High Flow” and “Received Signal” interruption
FILTER PROPERTIES
● 0.1, 10, 40 Hz: 2nd order Butterworth, with a
third passive pole at 160 Hz
● 160 Hz: 3rd order Butterworth
ZERO FLOW ADJUST
Recessed momentary push button to zero ow
reading at stopped ow.
FLOW OUTPUT
Front panel BNC output connector & rear panel
terminal block:
● Pulsatile/Average Volume Flow
● Filtering controlled by front panel selectable
lters
● Voltage range: -5 to + 5 volts
● Output resistance: 500 Ohm
● Full Range for Flow: -5 to +5 V (bidirectional
ows, with range of 5 times scale factor ow)
AUTOMATIC DIGITAL SENSOR ID & CALIBRATION
TS410 reads operational data (size, scale &
calibration) programmed in the Sensor’s EPROM.
ULTRASONIC FREQUENCY RANGE
600 KHz to 14.4 MHz; Sensor size dependent
SIGNAL OUTPUTS
8 accessible signals via 400-Series Console’s back-
panel terminal block: Pulsatile Volume Flow; Mean
Volume Flow; Received Signal Amplitude (2);
Phase (4)
SYNCHRONIZATION
Rear panel jumpers select synchronization mode
● Self-Triggering Mode: “SYNC IN” to “SYNC
OUT” jumper on each Module
● Sequential Triggering Mode: “SYNC IN” crossed
to “SYNC OUT” between Modules