EasyManua.ls Logo

Transonic TS420 - TS420 Chronic Application and Flow Measurement

Default Icon
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26 AU-OPR-400Ser-EN,
Rev D
i. SP430 Module: Functions & Controls
MEA (MEASURE) MODE BUTTON
Press to engage pulsatile pressure measurement mode. Light indicates activation.
0 MMHG CALIBRATION BUTTON
Outputs calibration voltage (-2.857 V ±2%) corresponding to 0 mmHg. Light indicates activation.
100 MMHG CALIBRATION BUTTON
Outputs calibration voltage (-0.571 V ±2%) corresponding to 100 mmHg. Light indicates activation.
HDMI CONNECTORS FOR PRESSURE INPUTS
Pressure Catheter extension cable or YS100-ADPT adaptor with extension cable. Two channels available.
Use only Transonic
®
approved HDMI cables.
BALANCE CONTROL KNOBS
Adjusts the pressure output voltage for nullifying zero offset. Each channel is adjusted separately.
NOTE: KNOBS ADJUST EASILY - BEWARE OF UNINTENDED ADJUSTMENTS.
BNC ANALOG OUTPUTS
Pulsatile pressure as a voltage. One output for each channel.
A. Front Panel
MEA (Measure) Mode Button
0 mmHg Calibration Button
100 mmHg Calibration Button
HDMI Connectors
Balance Control Knobs
BNC Analog Outputs
Fig. 5.4: SP430 Transonic Scisense
Pressure Amp Module
Indicator Lights

Table of Contents

Related product manuals