28 AU-OPR-400Ser-EN,
Rev D
ii. SP430 Module: Specications
GENERAL FEATURES
Weight/Size Dimensions: 5.125” h x 2” w x 9.062” d
Weight: 1.9 lbs
Module is 1 Console bay (10HP) wide
Console Compatibility: 400-Series Flowmeter Consoles (T402 and T403).
Power: Derives input power from 400-Series Console 12Vdc ±5% input voltage.
RoHS compliant
Installation in a Console is required. The SP430 module must be installed in one of the bays on the far right
in the T402 or T403 console (slot, 3, 4, 5 or 6) to meet specications for best performance. Location in slot 1
or 2 may result in excessive noise in the signal.
CATHETER/TRANSDUCER COMPATIBILITY
Accepts Pressure input in 2 front panel HDMI connectors. Each channel has separate controls and outputs.
Transonic Scisense solid-state Pressure Catheters with HDMI connectors
Transpac
®
IV from ICU Medical using the YS100-ADPT RJ11 TO HDMI adaptor
ANALOG OUTPUT SIGNALS
Front panel BNC: 2 analog pressure signals ltered to 800 Hz
Rear Panel Terminal Connector: 2 analog pressure signals ltered to 800 Hz;
Internal atmospheric pressure reference ltered to 1 Hz
ACCURACY
Refer to individual Pressure Transducer specications
INPUT FILTER
3.5 kHz low pass
OUTPUT FILTER
800 Hz low pass
AMPLIFIER GAIN
1143
NOISE
±10 mv or 1 mmHg
RANGE
-50 to + 300 mmHg (2.286 V/100 mmHg)
BAROMETRIC PRESSURE REFERENCE
Inneon KP236 Analog absolute Pressure Sensor
Accuracy: ±1 kPa.
Sensitivity: 53.3 mV/kPa
Operating range: 40-115 kPa