EasyManua.ls Logo

Transonic TS420 - TS420 Operation Modes and Filters; TS420 Analog Outputs and Calibration Keys

Default Icon
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AU-OPR-400Ser-EN,
Rev D
19
B. Modes of Operation
C. Filter Settings
To record instantaneous ow: set the ow
output [Filter] to a frequency at least 10 times
the ow pulsation or heart rate. The digital
sample rate should be set to a minimum of 3
times the application frequency.
D. Analog Outputs
Transonic
®
Flow Modules output analog signals in the range of -5 volts to +5 volts that are compatible
with most A/D (analog-to-digital) converters of Data Acquisition systems. The back panel analog outputs
generated by the TS420 Module are listed in the table below:
E. Calibration (Cal) Keys
The Calibration Keys store Probe specic calibration information normally located in the Probe’s EPROM.
Probes with small 4-pin connectors do not have a built in EPROM and require the matching Cal Key to
function. NOTE: Cal Keys are Probe specic and should not be used with Probes other than the one for
which they were programmed. Calibration keys will override a Probe connector EPROM if both are present.
NUMBER TERMINAL NAME VOLTAGE DESCRIPTION
12 Sync In N/A In/Out Jumper required for operation; Multi-Module synchronization input
11 Sync Out N/A In/Out Jumper required for operation; Multi-Module synchronization output
10 GND Ground
9 GND Ground
8 Rec Amp 1 0 - 4 V Quality of ultrasound transmission for testing Probe functionality, acoustic coupling or
blockage of ultrasound by air or other impedance mismatch. 2 V = 100%
7 Rec Amp 2 0 - 4 V Same as Rec Amp 1 for a second pair of transducers. 2 V = 100%
6 Phase 2 B ± 5 V Same as Phase 2 A but offset by +4.5 V or -4.5 V
5 Phase 2 A ± 5 V Same as Phase 1 A but for a second pair of transducers
4 Phase 1 B ± 5 V Same as Phase 1 A but offset by +4.5 V or -4.5 V
3 Phase 1 A ± 5 V Acoustic velocity of uid used in ultrasound indicator dilution studies
2 Mean Flow ± 5 V Average volume ow output, ltered at 0.1 Hz
1 Puls Flow ± 5 V Instantaneous pulsatile volume ow output, ltered at 10, 40, or 160 Hz depending on
the Module front panel [Filter] setting. Default to 160 Hz if 0.1 Hz lter is used.
Only PAU-Series Probes have two transducer pairs, all other Probe series have a single pair. For single pair Probes Phase 2 and Rec Amp
2 outputs are identical to Phase 1 and Rec Amp 1 outputs.
MODE
OUTPUTS
FRONT DIGITAL DISPLAY FRONT ANALOG DISPLAY
MEAN & PULSATILE FLOW
Front BNC/ Rear Analog
MEASURE
Bar indicator of signal quality
Average ow in mL/min or L/min
Start-up & error messages
Instantaneous volume ow= ow reading x
scale factor
Flow = recorded voltage x scale factor
Bidirectional ow output ± 5 V
TEST
Probe size and received signal
quality
Proportional value of received signal amplitude
1 V = standard normalized level
ZERO 0 mL/min 0 V = zero ow (calibration reference) 0 V = 0 mL/min
SCALE Scale factor ow value 1 V = scale factor ow (calibration reference) 1 V = scale factor ow
TS420 Module: Functions & Controls
HEART RATE OR
APPLICATION FREQUENCY
LOW PASS
FILTER
SETTING
RECOMMENDED MINIMUM
DIGITAL SAMPLE RATE FOR
DATA ACQUISITION
Average Flow Recording 0.1 Hz 0.3 Hz
Pulsatile to 60 beats/minute 10 Hz 30 Hz
Pulsatile to 240 beats/minute 40 Hz 120 Hz
Pulsatile to 960 beats/minute 160 Hz 500 Hz

Table of Contents

Related product manuals