iii. Consoles: Specications
WEIGHT/ SIZE
T402 Four Bay Console
5.21” h x 9.25” w x 12” d
5.8 lbs
T403 Six Bay Console
5.21” h x 13.46” w x 12” d
7.6 lbs
Consoles have side panel handles and tiltable front feet for easy viewing
MODULE COMPATIBILITY
Accepts 400-Series Flow and Pressure Amplier Modules
ELECTRICAL
Console is grounded. If accidentally left ungrounded, line to isolation ground leakage current is less than
50 microamperes.
RoHS COMPLIANT
POWER
AC Input: 100-240 VAC; 50-60 Hz, 50 watts
Fuses: 0.8A fast blo, mfg bussman # GMA0.8, 250 VAC
POWER CORD
USA/Japan: Feller 458-H161 or equivalent
Europe: Feller 199-000 or equivalent
United Kingdom: Feller 209-000 or equivalent
Australia: Feller 198-000 or equivalent
AUDIBLE ALARM
Beeping alarm, non-adjustable volume. Trip level and on/off setting activated through TS410 program.
MODULE CAPACITY
T402 Four Bay Console: Accepts 2 double-bay, wide Modules of 20 HP width or 4 single bay, narrow
Modules of 10 HP width or a combination of wide and narrow Modules.
T403 Six Bay Console: Accepts 3 double-bay, wide Modules of 20 HP width or 6 single bay, narrow
Modules of 10 HP width or a combination of wide and narrow Modules.
CONSOLE TO MODULE CONNECTION
96-pin DIN connector on proprietary backplane. Thumb-screw to lock into front panel.
SIGNAL OUTPUTS
Back panel screw terminal block receives output signals from Module(s) via DIN 96-pin connection with
Console; 12 output connections per Module, two dedicated for ground; See Module specications for
signal denition and voltage rating. Use with general purpose hookup wire; stranded wire is preferred.
Wire: UL 1007 or equivalent; 24 - 14 gauge
Minimum strip length: 6 mm (0.236 inches)