AU-OPR-400Ser-EN,
Rev D
35
Appendix B: EMC Tables
The 400-Series System complies to the Standard for Safety Electrical Measurement and Test Equipment Part
1: General Requirements UL 3111-1 and Safety Requirements for Measurement, Control and Laboratory Use
Part 1: General Requirements CSA C22.2 No. 1001.1-92 in accordance with UL 61010A-1 Electrical Equipment
for Laboratory Use, Part 1: General Requirements.
In accordance with EN 61326:1997 Electrical Equipment for Measurement, Control and Laboratory Use, the
following tables provide information regarding the EMC characteristics of this Electrical Equipment.
MANUFACTURER’S DECLARATION - ELECTROMAGNETIC EMISSIONS
The 400-Series System is intended for use in the electromagnetic environment specied below.
The customer or user of the 400-Series System should ensure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE
Radiated emissions
EN 61326
Class A
Conducted emissions
EN 61326
Class A
MANUFACTURER’S DECLARATION - IMMUNITY
The 400-Series System is intended for use in the electromagnetic environment specied below.
The customer or user of the 400-Series System should ensure that it is used in such an environment.
IMMUNITY TEST EN 61326 TEST LEVEL COMPLIANCE LEVEL
ESD
IEC 1000-4-2
±2kV Contact
±4kV Contact
±2kV Air
±4kV Air
±4kV Indirect via HCP
±4kV Indirect via VCP
Pass:
Performance criteria B
RF Electromagnetic Field
IEC 1000-4-3
80-1000 MHz, 3 V/m, 80%
AM (1kHz)
Pass:
Performance criteria A
BFT
IEC 1000-4-4
±0.5kV, 5/50 Tr/Th
±5kHz I/O
Pass:
Performance criteria B
EFT
EN 61000-4-4
±2kV Mains Pass:
Performance criteria B
Surge
IEC 1000-4-5
±0.5kV Differential
±1kV Common
Pass:
Performance criteria B
Conducted RF
IEC 1000-4-6
3 Vrms, 80% AM (1 kHz)
0.15-80 MHz to Main & I/O
Pass:
Performance criteria A
Voltage Dips/Dropout
IEC 1000-4-11
Reduction 100%, 1 Cycle
Pass:
Performance criteria B