The control unit prompts for the mixture Gmm. Gmm is the
maximum theoretical specific gravity.
Enter the Gmm value using the number keys. Press 〈ENTER〉.
The control unit displays the date and time of the last six data sets
stored in memory locations 1 through 6. Select the first data set by
pressing the number key that corresponds to the number of the
memory location. The control unit prompts for the weight of the
first sample. Enter the weight of the specimen using the number
keys. Press 〈ENTER〉. The control unit prompts for the Gmb for the
first specimen Gmb is the bulk specific gravity of the compacted
specimen. Enter the Gmb for the sample using the number keys.
Press 〈ENTER〉.
If the user chooses to average more than one data set, then the
control unit again displays the date and time of the last six data sets.
Repeat the selection for each data set you want to include in the
average. If the user chooses a data set with a different number of
gyrations, then the control unit will display an error. Select each
data set and enter the weight and actual Gmb entry for each set.
NOTE
When averaging data sets, all selected data sets must
have the same number of gyrations.
After entering the weight and Gmb for each data set, the control unit
displays the output options:
1
-
Print Table
2
-
Graph & Table
3
- Exit