EasyManua.ls Logo

Troxler 4140 - Page 64

Troxler 4140
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-20
Install a clean bearing.
Generously coat the top of the bearing with Magnalube-G.
Apply Magnalube-G to the inside diameter of the cavity.
Install a clean o-ring in the groove around the clean turntable.
Ensure the o-ring is fully seated in the groove.
Carefully insert the turntable into the cavity.
Ensure that the plate is fully seated and the o-ring is still
properly installed around the turntable.
Reinstall the clean retainer. Tighten the six 4-40 x 1/4" flathead
screws in a crisscross pattern.
NOTE
When compacting specimens at an angle greater than
0.7°, do not lubricate the turntable.

Related product manuals