3-20
Install a clean bearing.
Generously coat the top of the bearing with Magnalube-G.
Apply Magnalube-G to the inside diameter of the cavity.
Install a clean o-ring in the groove around the clean turntable.
Ensure the o-ring is fully seated in the groove.
Carefully insert the turntable into the cavity.
Ensure that the plate is fully seated and the o-ring is still
properly installed around the turntable.
Reinstall the clean retainer. Tighten the six 4-40 x 1/4" flathead
screws in a crisscross pattern.
NOTE
When compacting specimens at an angle greater than
0.7°, do not lubricate the turntable.