Install a clean bearing.
✓
Generously coat the top of the bearing with Magnalube-G.
✓
Apply Magnalube-G to the inside diameter of the cavity.
✓
Install a clean o-ring in the groove around the clean turntable.
Ensure the o-ring is fully seated in the groove.
✓
Carefully insert the turntable into the cavity.
✓
Ensure that the plate is fully seated and the o-ring is still
properly installed around the turntable.
✓
Reinstall the clean retainer. Tighten the six 4-40 x 1/4" flathead
screws in a crisscross pattern.
NOTE
When compacting specimens at an angle greater than
0.7°, do not lubricate the turntable.