EasyManua.ls Logo

Troxler 4140 - Page 82

Troxler 4140
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-4
CONTROL UNIT DISPLAYS: Printer Out of Paper Press
ENTER To Continue Or ESC To Abort
See the printer Operating Instructions for details on installing
paper.
CONTROL UNIT DISPLAYS: Printer Offline Press ENTER
To Continue Or ESC To Abort
Press ON LINE on the printer. See the printer Operating
Instructions for details on the printer control panel.
CONTROL UNIT DISPLAYS: No Data to Plot
Compact a specimen before attempting to plot.
CONTROL UNIT DISPLAYS: Serial Printer Not Ready
Check the serial cable and serial connection.
When using GyroPave software, select Start from the Load
Data From Compactor screen before trying to load data either
automatically or manually.
HEIGHT-VERSUS-GYRATIONS DATA DOES NOT
DOWNLOAD TO THE COMPUTER
Check that the Auto Output feature is turned on (see page 2-12).
Check that the serial port is specified for the Auto Output
feature (see page 2-12).
Check serial parameters: 9600 baud rate, 8 data bits, 2 stop bits,
and no parity.
When using the GyroPave software, select Start from the Load
Data From Compactor screen before trying to load data either
automatically or manually.

Related product manuals