EasyManuals Logo

Troxler ICO 4740 User Manual

Troxler ICO 4740
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
ICO 6-7
CALCULATING AN ACF
1. Using the following equation, calculate and record the
aggregate correction factor for each correction sample:
ACF =%AC
0 %ACT
where: ACF = Aggregate correction factor
%AC
0 = Asphalt content of sample as determined
by the oven
%ACT = True asphalt content
2. Determine the average ACF value
(ACFAVG)
as follows:
If the difference between the two ACF values is less
than 0.15%, then average the two values.
If the difference between the two ACF values is
greater than 0.15%, then prepare two additional
correction samples. Discard the high and low ACF
values, and then average the two remaining values.
3. Enter the ACF as described in the Create ACF section on
page 6-9. The oven will automatically adjust the asphalt
content measurements using the active ACF.

Table of Contents