ICO C-3
PERFORMANCE SPECIFICATIONS
Maximum sample size
2500 g per sample basket
5000 g total
Integrated scale resolution
Burn time for 1200 g
Internal memory capacity:
Sample data
Aggregate correction
factors
Project IDs
Temperature sensors
Type K thermocouples
(see note below)
NOTE
The ignition method employed by the ICO does not
require precise control of the chamber or other
temperatures. If desired, oven performance can be
verified using a precision asphalt mix.
The oven thermocouples are calibrated at the factory
and do not require re-calibration by the operator. If
verification of the thermocouples is desired, Troxler
recommends checking their accuracy at room
(ambient) temperature.