ICO 4-5
FIRST-TIME STARTUP AND SETUP
WARNING
To prevent personal injury or equipment damage, the
operator should become familiar with the safety
warnings and information in Chapter 2 before
operating the ICO.
This section summarizes the initial startup and setup of the
oven. For more detailed instructions, see Chapters 5 through 7,
as referenced in this summary.
1. Ensure that the oven is installed as described in Chapter
3.
2. With the chamber door open, position the circuit
breaker/power switch, found in the lower left of the
oven rear panel (see Figure 1-1), to the ON (up)
position. This breaker does not need to be turned off
during normal operation. Press the on/off rocker
switch on the right-hand side of the front panel to the on
position (1 for on, 0 for off). The oven performs a self-
test, and then displays the following default screen. The
second line of the default screen shows the current
chamber temperature (25
°
C in this example). Note that
the open chamber door prevents the heating element
from heating. The chamber must be at room
temperature to verify the scale as described in step 4.
Proj PROJECT NAME
Temperature 25 C
Mass 0.0 g
11/17/2001 8:06 AM