ICO 5-27
STORE FUNCTION
After burning a sample and before burning another sample,
sample data can be stored under the active project ID. (See
Chapter 7 for details on creating a project ID.) The Store
selection from the MAIN MENU enables the operator to store
sample data manually.
NOTE
This function can also be accessed directly from the
default screen by pressing the (STORE) key.
After burning a sample and before burning another sample,
press
(7) at the MAIN MENU or press the (STORE) key if the
default screen is displayed.
If there is no sample data to be stored, the oven displays an
error message and returns to the previous display.
NOTE
The Auto-Store sample ID number is not incremented
if sample data is stored manually. When using the
manual Store function, the operator enters a sample
ID number manually. Therefore, duplicate sample ID
numbers could be stored under the active project.
For example, if the Auto-Store function is turned on
for a new project, the oven stores the sample data for
the first two burn cycles under sample IDs 1 and 2.
Assume that the Auto-Store is then turned off, and the
operator manually stores data from two more burn
cycles using sample IDs 3 and 4. If the Auto-Store
function is then turned on, the sample data for
successive burn cycles will also be stored using sample
IDs 3, 4, and so on.