Operation
21
Analog Output
Newer models of the Model 8520 D
USTTRAK Aerosol Monitor contain an
analog output feature. This means the instrument is capable of providing an
analog voltage signal that is proportional to mass concentration.
Note: Before using this feature, it is important to read this operational
section. For further information, please see Application Note
ITI-073, which describes the Analog Output programming and
operation. This Application Note is available at the TSI website,
http://www.tsi.com.
Because the D
USTTRAK monitor measures aerosol over such a wide dynamic
range, it is necessary to select the appropriate range before using the analog
output feature. The instrument may be programmed to one of four options,
using the T
RAKPRO software. When shipped from the factory, the instrument
is programmed to the lowest range. To select the analog output range, do the
following:
1. Turn on the D
USTTRAK monitor and start the TRAKPRO software
(version 3.2 or higher).
2. Select P
arameters, Analog Output, from the Instrument Setup menu.
The following dialog is displayed:
3.
Select one of the four ranges, and press Send.
Please note the following considerations when using the analog output
feature.
The analog output function is always “On.” You do not need to activate
this function.
The interpretation of the analog output voltage is directly tied to the
chosen range. For example, at the lowest range, each “0.001 mg/m
3
” of