49
Appendix A
Specifications
Specifications are subject to change without notice.
Sensitivity:
Sensor type: ..................... 90° light scattering, laser diode
Range: ............................. 0.001100 mg/m
3
(calibrated to respirable
fraction of standard ISO 12103-1, A1 test
dust)
Resolution: ...................... ±0.1 % of reading or ±0.001 mg/m
3
,
whichever is greater
Zero stability: .................... ±0.001 mg/m
3
over 24 hours using
10-second time-constant
Particle size range: .......... 0.1approximate 10 µm (Upper limit is
dependent on flow rate)
Flow rate range: ............... 1.42.4 L/min.
Temperature coefficient: .. +0.001 mg/m
3
per °C (for variations from
temperature at which D
USTTRAK monitor
was zeroed)
Instrument Temperature Range:
Operating range: ............. 0
°C to 50 °C (32 °F to 120 °F)
Storage range: ................ -20
°C to 60 °C (-4 °F to 140 °F)
Time Constant:
Range: .............................. Adjustable from 1 to 60 seconds
Data Logging:
Data points: ...................... >31,000 (21 days logging every minute)
Logging interval: ............... Adjustable from 1 second to 1 hour
Analog Output (selected instruments only):
Selectable ranges: ........... 0 0.100 mg/m
3
0 1.0 mg/m
3
0 10.0 mg/m
3
0 100 mg/m
3
Voltage output .................. 0 5 VDC
Output impedance ............ 0.01 ohms
Maximum current ............. 15 mA