45
Chapter 5
Troubleshooting
The table below lists the symptoms, possible causes, and recommended
solutions for common problems encountered with the D
USTTRAK monitor.
Symptom Possible Cause Corrective Action
reading.
Dirty inlet port and/or
sample tube.
Sample nozzle missing.
Internal filter(s) not
installed properly
(leaking).
Check connections for leaks.
Replace zero filter.
Tighten inlet cap.
Clean inlet port. Clean or
replace tubing.
Remove inlet port cap and
check sample nozzle. Replace
if missing.
Inspect internal filter wells to
make certain the filters and o-
rings are seated properly.
Replace internal filters if
No display. Unit not switched on.
Low or dead batteries.
Batteries installed
backwards.
Dirty battery contacts.
Switch unit on.
Replace the batteries or plug in
the AC adapter.
Observe polarity indication on
battery holder diagram.
Clean the battery contacts.
No keypad
Display/Lockout switch
Slide lockout switch to normal
Incorrect
function
displayed.
Two keys have been
pressed at the same time.
Press only one key at a time.