Chapter 3
2. Select Logging Setup from the Instrument Setup menu. The TRAKPRO
software retrieves the current settings for LOG 2 and LOG 3 modes
from the D
USTTRAK monitor and displays them in the previous dialog.
3.
Enter the following for LOG 2 and LOG 3:
4. Select S
end. The logging instrument is programmed for the mode 2 and
mode 3 protocols.
5. Note that the LOG 2 test requires 1% of the available memory and
LOG 3 requires less than 1% of the memory. A total of 100% of the
memory is available for use.
6. You can now disconnect your D
USTTRAK monitor from the computer.
Refer to other sections of this Operation and Service Manual for details
on making measurements using LOG 2 and LOG 3 modes.
After programming the D
USTTRAK monitor with the TRAKPRO Data
Analysis Software, take the D
USTTRAK monitor to the desired location and
turn it on. Put it into LOG 2 or LOG 3 mode (whichever you programmed)
using the SAMPLING MODE key. Press the SAMPLE key to initiate the
program.
If you have set a start time and/or date, the display toggles between the next
“TEST ID” and the “ELAPSED TIME 0” message. If the test start time is
greater than one minute away, the D
USTTRAK monitor shuts off until one
minute before the test start time. This indicates that the D
USTTRAK monitor
is waiting until the programmed start time and dates occur. To prevent
tampering, lockout the display and keypad at this time.
Things You Should Know About Taking Pre-Programmed
Samples
If you press the SAMPLE key during programmed operation, the
program terminates (unless the keypad is locked).