Troubleshooting the Environmental Enclosure
Symptom Possible Cause Corrective Action
Readings are
unusually
low (cont.).
Leak in the inlet.
(cont.)
Make sure inlet assembly
O-ring is in place.
Re-grease the O-rings sealing
the inlet assembly and the
water trap bottle.
DUSTTRAK
monitor
Display
reads
“MEMORY
Memory is full. Download memory to a PC if
desired then clear memory.