Page 31 of 58 Pages
previous cycle, a countdown of time left until the 10 minutes
elapse will be displayed in the interval between the readouts
of temperature and pressure.
When a cycle is started by means of pressing the START
key, the load number is displayed for 2 seconds on the left of
the lower row.
Examples
Example 1:
Autoclave between processes, the program No.1 has
been selected.
Example 2:
The autoclave in the sterilization stage, program No.3 is
running. The time left to completion of 3 minutes and
14 seconds.
Example 3:
The process failed due to temperature drop in the
sterilization stage in program No.2.
2 - Instruments EXH.
LOW TEMP. 178K
1 - Instruments ST.BY
35
C 100K
Temperature Timer Pressure
Selected program Waste Stage
3 - Waste STER
122ºC 03:14 212K