Low Water Message is displayed, FAIL indicator lights and the program s
aborted.
If the electrode sensor indicates no water and the safety
thermostat cuts-off during the heating stage.
If the thermostat cuts off, and the micro-controller
presumes the electrode sensor is damaged.
If a power failure occurred during the HEAT or STE.
stage, the cycle will resume when power returns, provided
there is enough water in the chamber. If not the cycle will
abort, the message LOW WATER displayed, and the red
FAIL indicator lights up.
Electrode Message is displayed if the water level electrode is dirty. In
this case the message “clean electrode will be printed at the
end of the cycle on the printer’s output.
No Liquids! Message is displayed, on the bottom row, after pressing the
START/ STOP key to start cycles 1 or 2 (Instruments). This
massage is intended to remind the operator not to sterilize
liquids with these two programs. The message disappears after
pressing again the START/ STOP key.