PRINTER OUTPUT DESCRIPTION
Autoclave:1 Number of autoclave.
Operator :___________ To be filled in manually by operator.
05/01/2010 10:14:20 Date and time sterilization cycle ended.
Load number: 0002 Load number. Useful to determine when to clean the chamber.
Cycle ended
E30:35 118.1°C 105.0k The time, temperature and pressure during exhaust.
-----------------------
S28:23 135.0°C 309.6k The time, temperature and pressure during sterilization.
*
* Prints sterilization data every 1 minute.
*
S19:23 134.9°C 310.6k The time, temperature and pressure during sterilization.
S18:23 134.2°C 307.8k The time, temperature and pressure during sterilization.
----------------------
H16:08 125.2°C 243.8k The time, temperature and pressure during heating.
----------------------
H01:08 020.5°C 097.8k The time, temperature and pressure during heating.
----------------------
W00:00 020.6°C 098.0k The time, temperature and pressure during water inlet.
01/01/2000 05:14:56 Date and time sterilization cycle begun.
Dry time: 000min Drying time for selected program (not applicable in these
models.
Ster time: 010min Sterilization time for selected program.
Ster Temp: 134°C Sterilization temperature in chamber for selected program.
Cyc: 1-Instruments Selected program: 1 Unwrapped Instruments 134 cycle
Ver = Lab00Vn1 Number and version of the program
Legend
W