8.3.2 Loading
Correct loading of the autoclave is essential to successful sterilizing
for several reasons. Efficient air removal from the chamber and the
load will permit effective steam penetration and saturation, and
allow proper drainage of condensate. Additionally, correct loading
will prevent damage to packs and their contents and maximize
efficient use of the sterilizer.
For detailed loading instructions, see para. 6 (Preparation before
sterilization)
8.3.3 Unloading
On completion of the cycle, take out the load immediately from the
sterilizer. Do not remove the load from the basket until its
temperature reduces to the room temperature. Let the load cool
down in an area without air movement (air conditioning, etc.) and
with minimum people passing by to avoid possibility of touching
the hot load. Do not touch the hot load since hot load absorbs
moister and, therefor, may absorb bacteria from your hand. Do not
transfer hot load to metal shelves for cooling. Perform a visual
inspection to ascertain that sterilizing indicators have made the
required colour change, and that the load is dry.
The load shall be rejected if:
a. The package has been compressed.
b. The package is torn.
c. The load is suspected to be wet.
d. The load fell on the floor.
e. Condensed drops can be detected on the load.
To avoid injuries use heat resistant gloves while unloading the
autoclave.