4.2
FILLING
THE
WATER
RESERVOIR
1.
Remove
the
water
reservoir
cover.
Pour
distilled
water
into
the
reservoir
through
the
opening
on
top
of
the
autoclave
until
it
reaches
the
base
of
the
safety
valve
holder.
It
is
recommended
to
use
water
having
the
following
characteristics:
1.
Electrical
conductibility
lower
than
15
microsiemens.
2.
Chlorides
content,
less
than
4
pp.m.
3.
PH
6.5-8.
Caution:
Under
no
circumstance
should
water
be
filled
above
the
safety
valve
holder.
Use
distilled
water
only
.
Tap
water
may
block
the
hole
of
the
air
trap.
This
causes
air
pockets
which
prevent
the
temperature
from
rising.
It
is
essential
that
from
time
to
time
during
the
sterilization
period
,
a
spray
of
steam
should
escape,
causing
a
hissing
sound.
If
no
steam
is
evident,
follow
instructions
for
air
trap
jet
cleaning
procedure
on
page
30.