OPERATING
INSTRUCTIONS
Insert
the
plug
into
the
socket.
Remove
water
reservoir
cover.
Pour
distilled
water
into
the
reservoir,
through
the
opening
on
top
of
the
autoclave,
until
it
reaches
the
base
of
the
safety
valve
holder,
approximately
0.7
gallons
(3
liters).
Turn
on
the
rocker
switch
mounted
on
the
bottom
of
the
front
panel
to
power
the
control
circuits.
Set
the
clock
for
the
proper
date
and
time.
Press
the
key
to
select
the
required
program
.The
light
indicator
of
the
given
program
is
illuminated,
indicating
the
program
has
been
selected.
The
present
data
of
the
program,
Sterilization
temperature,
time
and
dry
time
are
displayed.
Load
the
material
to
be
sterilized
into
the
chamber,
close
the
door
making
sure
the
Close
Door
indicator
is
illuminated.
Press
the
START
key
to
put
the
autoclave
in
operation.
WATER
INLET
is
displayed
until
the
correct
volume
of
water
is
automatically
introduced.
The
autoclave
starts
performing
the
sequence
of
operations.
The
actual
measured
values
of
pressure
and
temperature
are
displayed
continuously
and
printed
every
minute
(with
optional
printer).
The
indicator
lights
HEAT,
STE,
and
DRY
are
lit
as
explained
before.
NOTE
:
It
is
possible
to
change
parameters
1-7
Page12,
only
when
the
autoclave
is
not
in
operation.
In
order
to
change
the
fixed
preset
parameters,
proceed
as
follows:
To
increase
or
decrease
the
sterilization
temperature,
sterilization
time
or
dry
time,
follow
instructions
from
the
chapter
1.4
.
Pressing
the
UP/
DN
key,
advances
the
setting
upwards
or
downwards
by
one
unit.
The
displays
are
updated
with
every
change
in
the
preset
data.
During
program
2,
the
dry
stage
begins
after
the
steam
exhaust
stage,
The
exhaust
valve
is
open,
while
the
chamber
is
heated,
which
allows
evacuation
of
water
vapors
and
humidity
from
the
chamber.
For
models
EA
or
EKA,
an
air
compressor
which
works
during
the
dry
stage
pushes
hot
air
which
flows
through
a
0.2
yu
filter
enters
the
chamber
and
escapes
through
the
opposite
end
exhaling
the
water
vapors
and
improving
drying
of
the
sterilized
material.