25.
EKA
ONLY
-
Pump
and
valve
always
ON
or
OFF
e
Check
the
S.S.R2
in
the
electric
box.
e
Check
if
TP
18
is
in
0.0V
if
on
and
5.0V
if
off,
if
yes
replace
PREDG.
If
not
replace
the
AJUNC3
board.
26.
EKA
-
Pump
is
always
on,
valve
is
off
e
The
problem
is
in
the
pump
or
in
SSR2
located
in
the
electric
box.
7.2
HOW
TO
CHANGE
THE
CARTRIDGE
FUSE
Caution
ZN
Make
sure,
the
electrical
power
cord
is
disconnected!
Use
a
screwdriver
to
unlock
the
fuseholder
cover
by
turning
it
counter
clockwise
a
little,
and
pull
it
out.
Insert
a
new
Cartridge
into
the
holder
and
turn
the
cover
clockwise
until
locked.
Ensure
the
correct
fuse
is
installed
as
marked!
The
bottom
two
protects
line
and
Neutral
Fuse
ISA
type
T5
x
20
or
Fuse
10A
type
T5
x
20
The
top
one
protects
the
air
pump
Fuse
1.2
A
type
T
5x20.
39