20.
LOW
PRESS
MESSAGE
IS
DISPLAYED
This
is
usually
due
to
insufficient
water
inside
the
chamber.
e
If
the
water
valve
is
OK,
the
electrode
detecting
the
water
level
may
be
damaged.
It
may
detect
water
in
the
chamber
although
it
is
actually
empty.
e
The
normal
operation
procedure
of the
electrode
is
to
let
water
in
the
chamber
8
additional
seconds
after
it
has
detected
a
sufficient
water
level.
If
the
electrode
is
short
circuited,
it
may
let
water
into
the
chamber
for
ONLY
seconds.
e
Check
if
there
is
any
contact
between
the
electrode
and
the
chamber.
Use
an
Ohm-
meter
for
that
purpose.
Make
sure
none
of
the
tools
to
be
sterilized
has
any
contact
with
the
electrode.
If
the
problem
persists,
the
electrode
circuit
on
the
AJUNC3
board
may
be
short
circuited
or
damaged.
21.
CLOCK
AND
DATE
ARE
INACCURATE,
NEW
DATA
IS
NOT
KEPT
IN
MEMORY
e
If
new
data
(operator’s
own
parameters
)
is
not
kept
in
memory,
then
the
backup
battery
is
down,
and
the
same
unit
has
to
be
replaced
as
the
backup
battery
is
an
integral
part
of
the
Real
Time
Clock.
Change
the
Real
Time
Clock
(U2),
on
PREDG.
e
The
battery
manufacturer
offers
a
ten
year
guarantee,
but
it
is
recommended
to
change
the
Real
Time
Clock
including
the
integrated
backup
battery
once
every
8
years.
22.
UNPREDICTED
FUNCTION
OF
THE
AUTOCLAVE
e
Check
the
input
and
output
voltages.
Disconnect
the
printer
as
it
may
be
causing
a
short
circuit
if
it
is
faulty.
If
the
problem
persists,
the
keypad
may
be
short
circuited,
replace
it
with
the
PREDG
board,
if
that
does
not
help,
replace
the
complete
electronic
system.
23.
DISPLAYING
OF
LOW
TEMPERATURE
If
LOW
TEMPERATURE
is
displayed,
the
following
should
be
checked
:
1.
The
proper
amount
of
water
is
in
the
chamber.
The
safety
thermostat
cuts
off
the
heating
elements
during
the
sterilization
stage.
To
raise
the
working
temperature,
do
one
of
the
following:
Using
a
plier
or
a
screwdriver,
turn
the
thermostat
stem
clockwise
1/8
-
1/4”.
Repeat
this
procedure
if
necessary.
24.
MEMORY
RESET
To
reset
the
memory
of
the
autoclave
control
unit
backed
up
by
a
battery,
proceed
as
follows:
Be
sure
that
there
is
no
pressure
in
the
autoclave
chamber.
Turn
the
main
switch
to
OFF
position.
Press
the
STOP
key.
At
the
same
time
turn
the
main
switch
to
ON
position.
Keep
pressing
the
STOP
key
until
the
program
parameters
are
displayed.
Set
the
Sterilization
temperature,
Sterilization
time,
Date
and
Time.
Repeat
the
procedure
of
automatic
water
filling
-
Page
20.