7.7
CUT-OUT
THERMOSTAT
This
thermostat
cuts
out
the
power
to
the
autoclave,
in
the
event
that
all
other
safety
systems
do
not
function.
For
example:
if
the
operator
foryets
to
fill
the
chamber
with
water,
and
then
starts
the
sterilization
cycle,
the
safety
thermostat
cuts
the
electricity
to
the
heaters.
If
it
does
not
and
the
temperature
still
rises,
then
the
cut-out
thermostat
will
cut-out
the
power
to
the
autoclave.
In
order
to
restart
the
operation,
press
the
Reset
Button.
If
the
autoclave
is
operated
according
to
the
instructions,
and
the
thermostat
again
cuts
out,
a
service
technician
should
be
called.
The
thermostat
can
be
calibrated
with
a
screw
driver
by
turning
the
center
adjust
screw
(1):
e
clockwise
to
raise
the
temperature.
e
counter-clockwise
to
lower
the
temperature.