USER'S GUIDE____________________________________________________________________
106 _________________________________________________________________ M210482EN-D
28), disconnect the blower cable from connector J1 (see Figure 11
on page 34) and pull out the unit.
3. Loosen the 12 screws on the fram
e of the window and remove the
window by lifting it out with the screws attached. Also remove any
pieces of the old gasket.
4. Use some solvent to clean the surface of the measurement unit
before attaching the new window. Additionally, check that the
gasket on the window is undamaged and in place.
NOTE
Avoid touching the new window with your bare hands as it may get
scratched. Avoid also touching the clean window surface as it may get
contaminated.
0910-113
Figure 23 Window CLW311
5. Place the window on the measurement unit and attach the 12
screws with your fingers to prevent the window from sliding off its
place.
6. Tighten the screws in a crosswise order to allow the window gasket
to tighten evenly. For example, you can first tighten the screw on
top of the window frame, then the one on the bottom, then the one
on the left, and the one on the right. Tighten the remaining eight
screws in a clockwise order, in such a way that you first tighten one
screw and then the one opposite to this, then the one next to the
first screw and so forth. After tightening these eight screws,
retighten the first four screws.
NOTE
Remember to use a thread-locking compound to the window screws
before final tightening.