USER'S GUIDE____________________________________________________________________
80 __________________________________________________________________ M210482EN-D
b22 (0040 0000) Battery low (W)
b21 (0020 0000) Transmitter expire warning (W)
b20 (0010 0000) Heater or humidity sensor failure (W)
C: b19 (0008 0000) High radiance warning together with b02 (W)
b18 (0004 0000) Engine, receiver, or laser monitor failure
warning (W)
b17 (0002 0000) Relative Humidity is high > 85 % (option)
(W)
b16 (0001 0000) Light path obstruction or receiver saturation
(also receiver failure active, b29) (A)
B: b15 (0000 8000) Blower failure (W)
b14 (0000 4000) (spare) (W)
b13 (0000 2000) (spare) (W)
b12 (0000 1000) (spare) (W)
A: b11 (0000 0800) Blower is ON
b10 (0000 0400) Blower heater is ON
b09 (0000 0200) Internal heater is ON
b08 (0000 0100) Units are METERS if ON, else FEET
9: b07 (0000 0080) Polling mode is ON
b06 (0000 0040) Working from battery
b05 (0000 0020) Always 0 (Single sequence mode is. N/A)
b04 (0000 0010) Always 0 (Manual settings are effective. N/A)
8: b03 (0000 0008) Tilt angle is > 45 degrees (W)
b02 (0000 0004) High radiance warning together with b19. (W)
b01 (0000 0002) Manual blower control
b00 (0000 0001) (spare)
3RD LINE
= End-of-Text and Carriage Return + Line Feed
CT25K Data Message No. 6
Message number 6 is similar to message number 1 but extended with a
sky condition line. For an interpretation, see section Sky Condition
Algorithm on page 48.
The following is an example of me
ssage number 6 format:
CTA2060 1st line 11 char.
30 01230 12340 23450 FEDCBA98 2nd line 31 char.
3 055 5 170 0 /// 0 /// 3rd line 30 char.
4th line 3 char.
-------------------
total 75 characters