Table 16 Error Group 4 (Byte 86)
Error Code Description
0 Transmitter OK
1 Transmitter expires (laser power low) (W)
2 Transmitter failure
3 Not used
4 Not used
5 Not used
6 Transmitter shutoff (laser temperature too high) (A)
Table 17 Error Group 5 (Byte 87)
Error Code Description
0 Status OK
1 Laser monitor failure (W) or
Blower failure (W) or
High radiation warning (W) or
Engine warning (W) or
Tilt angle warning (angle is > 45 degrees) (W) or
Battery voltage low (W) or
Battery failure (W) or
Humidity high (option) (W) or
Humidity Sensor failure (option) (W)
2 Not used
3 Memory failure (A)
4 Not used
Table 18 Error Group 6 (Byte 88)
Error Code Description
0 Temperature regulation OK
1 Heater failure (W)
Table 19 Error Group 7 (Byte 89)
Error Code Description
0 Always 0 (Data transmission. N/A)
1 Not used
2 Not used
3 Not used
4 Not used
5 Not used
6 Not used
7 Not used
8 Not used