Syntax Description
Example (shows 0 as current address, enter 5 as the new address):
addr
Address : 0 ? 5
Table 23 Seri Command
Syntax Description
seri<cr> Show current serial line settings.
seri [baud p d s]<cr> Set new serial line settings for RS-485 line (also
aects the service port). The new settings are
taken into use when the device is reset or
powered up.
baud = baud rate (300, 600, 1200, 2400, 4800,
9600, 19200, 38400, or 57600)
1)
p = parity
• n = none
• e = even
• o = odd
d = data bits (7 or 8)
s = stop bits (1 or 2)
Example (show current serial line settings):
seri
Baud P D S : 19200 N 8 1
1) Modbus output works only with baud rate 9600 and above.
Keep the serial line settings at default (19200 N 8 1) for compatibility with the
RDP100 display panel.
Table 24 Sdelay Command
Syntax Description
sdelay<cr> Show serial line delay (response time).
sdelay [ddd]<cr> Set serial line delay (response time).
ddd = delay, range 0 ... 255. Corresponds to 0 ...
1020 milliseconds.
HMDW110 User Guide M211726EN-D