Syntax Description
Example (shows default state without user adjustments - oset is 0 and gain is 1 for both humidity
(Cp oset and gain) and temperature):
l
Cp offset : 0.00000000E+00
Cp gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Table 38 Li Command
Syntax Description
li<cr> Enter values for oset and gain parameters for
user adjustment. Useful for restoring some
earlier state of user adjustment.
Use this command only to restore values you
have previously written down based on the
output from the l command, or to restore the
default oset and gain.
Example (shows adjustment has been applied to humidity measurement oset, overwrites it with
0):
li
Cp offset : -1.50922060E-01 ? 0
Cp gain : 1.00000000E+00 ?
T offset : 0.00000000E+00 ?
T gain : 1.00000000E+00 ?
Chapter 6 – Using Service Port With Terminal Application
51