Chapter 5  ________________________________________________________________ Operation 
 
VAISALA _______________________________________________________________________ 69 
Hood Heaters 
By default 0, no hood heaters are used. The hood heater option is factory 
installed and may be used or disabled in the CONF session. All the 
PWD22/52 heaters are switched ON below 2 °C and off at 5 °C. When 
using hood heaters, a separate 24 V heating power must be supplied. 
Hood heating power is 30 W for each hood. 
BAUD 
Use the following command to change the baud rate and communication 
type: 
BAUD value communication_type 
Baud rates 300, 1200, 2400, 4800, 9600 
Communication types E (7E1) and N (8N1) 
The new value is saved in EEPROM and it is used also after reset or 
power up. The default baud rate set at the factory is 9600 bps (7E1). The 
communication type definition is optional. It does not change if the baud 
rate is changed. 
The BAUD command displays the current baud rate and communication 
type. 
BAUD RATE: 9600 E71 
Analog Output  
Analog output is wired in the default configuration.  
The analog output connection is set at the factory as shown in Figure 17. 
You must do the differential measurement connection at the user's end, 
which is also shown in Figure 17.