5.  Check that PWD22/52 has power by opening the PWC22/52 
electronics enclosure cover and checking that the Run and 
Signal/Offset LEDs are blinking. 
6.  If no LED is on: 
-  Check the power supply voltage. 
-  Check that all the connectors are properly inserted. 
-  Check the power cable and connections. 
7.  If the Run LED is blinking (once per second for correct operation): 
-  Connect a maintenance terminal to the RS port. 
-  Check steps 1 and 2. 
-  Try resetting by turning the power off/on.  
8.  If the Run LED is lit: 
-  Try resetting as above. 
-  If the state is continues, it is probably caused by either the 
program memory or CPU fault. 
Message Exists but Visibility Value Does 
Not 
1.  PWD22/52 control electronics is probably working. 
-  Check the status information with the STA command (see 
section STA on page 75). If there are any active alarms, the 
visibility value does not exist. 
-  Check P12, M12, BACKSCATTER and LEDI. See Appendix 
A, Internal Monitoring, on page 105 for internal monitoring 
values of the limits. 
2.  Go to the PWD22/52 site. 
-  Check the cable connectors. 
-  Check the receiver and transmitter units. Follow the instructions 
in section Removing and Replacing Optical Units on page 88 
when removing the units. 
-  Visually inspect the condition of the electronics.