If warnings or errors are detected, one or several of the following texts
appear at the end of the message.
See Table 15 and Table 16 below.
Table 15 Hardware Error Texts
Receiver or transmitter contamination signal has
increased more than the ALARM limit given in the
LEDI signal is more than 7 V or less than -8 V.
Receiver/transmitter power is less than 10 V or
more than 14 V.
Offset frequency is <80 or over 170 (PWC22/52
hardware).
Signal frequency + offset frequency = 0, Signal
frequency - offset frequency < -1
Too low signal in receiver backscatter
measurement.
Error in RAM read/write check.
Measurement is out of limits.
RAINCAP
®
signal is close to zero.
PWL111 signal is out of limits.
Table 16 Warnings
Receiver or transmitter contamination signal has
increased above the WARNING limit selected in the
Transmitter Intensity
Low
LEDI signal less than -6 V.
AMBL signal is less than -9 V.
Visibility Not Calibrated
Visibility calibration coefficient has not been
changed from the default value (see section CAL
CAL
Use the CAL command to calibrate the visibility measurement. The
calibration is done using opaque glass plates with known scatter
properties. The plates belong to the PWA12 calibrator kit.
The command is the following:
CAL Calibrator_signal_value
For example:
>CAL 485