Mode 1
Mode 1 is the logarithmic mode and it uses ln (visibility) values for the
output. The calculation is the following:
( )
,
min coeffrange
out
XII
I
⋅+=
Current that flows to analog output current sink.
Specified lowest output current value (for example 4 mA)
range
Current range specified
(for example I
range
= 20 mA - 4 mA =16 mA)
,
ln
ln
)ln()ln(
)ln()ln(
min
max
min
minmax
min
=
−
−
=
VIS
VIS
VIS
VIS
VISVIS
VISVIS
X
i
i
coeff
Initial visibility value in meters
min
Lowest visibility value specified in configuration
(for example 10 m or 100 m).
max
Corresponding maximum visibility value (for example
2000 m or 20000 m).
The corresponding DAC output LSB value is calculated from I
out
by
using DAC scale parameters.
Mode 2
Mode 2 is the extinction mode, which gives lower output for higher
visibility. The maximum visibility must be higher than the minimum.
The corresponding currents are automatically handled so that the lower
current corresponds to the maximum visibility. The calculation is the
following:
( )
,
max coeffrange
out
XII
I
⋅−=
Current that flows to analog output current sink.
Specified highest output current value (for example 20 mA).
range
Current range specified
(for example I
range
= 20 mA - 4 mA =16 mA)