1. Introduction
PaX-i Plus / Insight Installation Manual 5
Cautions
It is critical that installers read and understand the installation
instructions fully before installation.
The installer must confirm that the system is installed
according to the instructions provided by this manual and
perform the appropriate procedures therein.
If the equipment has been stored at temperatures of below
10℃ (50℉) for more than a couple of hours, allow the
equipment to reach room temperature before applying mains
voltage.
Installation and related work must only be performed by
people authorized by VATECH.
Do not connect any items or equipment to this system which
are not part of the system: IEC60601-1-1 (3
rd
edition: 2005).
Any equipment not approved by VATECH must comply with
the applicable standards: IEC 60950-1 (2
nd
edition: 2005) for
IT equipment (Ex: PC) and IEC 60601-1 (3
rd
edition: 2005) for
medical electrical equipment.
All operators of this equipment are responsible for ensuring
that the requirements outlined in IEC 60601-1-1 (3
rd
edition:
2005): Safety Requirements for Medical Electrical Equipment
are fully met to ensure the safety of patients, operators and
the environment.
Never touch sensitive areas such as sensors during
installation. These areas are indicated at the applicable stages
during the installation procedures.
Use of wireless phones may interfere with the operation of this
equipment.
Use an ESD (electrostatic sensitive device) wrist band during
installation and connect it to a ground wire.
Touch a ground point to discharge static electricity before
handling PCB boards.